Frequency Parameter Policy

QDash frequency parameters distinguish exploration results, calibrated frequencies, and the drive frequencies used for an execution.

This is the agreed design policy for parameter updates, Qubex YAML export, and execution UX. It is not a description of fully implemented behavior; the implementation gaps are recorded below. Input resolution and snapshot semantics follow Parameter Resolution.

Parameter roles

ParameterMeaningIndependent persistent QDash parameter
coarse_qubit_frequencyExploration estimate used to start or repeat qubit calibrationYes
qubit_frequencyCalibrated qubit frequency obtained from Chevron or RamseyYes
control_frequencyDefault control drive frequency resolved from qubit_frequencyNo
resonator_frequencyResonator frequency estimated by an experimentYes
optimal_readout_frequencyFrequency selected by readout optimizationYes
readout_frequencyDefault readout drive frequency resolved from the readout parametersNo

The persistent values are the accepted results used by subsequent tasks. Task history retains measurements regardless of whether they are accepted. A newer history record does not necessarily replace a persistent value.

coarse_qubit_frequency is an exploration input, not simply a less precise alias for qubit_frequency. Keeping the values separate allows spectroscopy to be repeated without replacing a working calibration. They may differ normally. Chevron and Ramsey do not copy their results back into the exploration parameter.

resonator_frequency and optimal_readout_frequency describe different measurement objectives. Readout optimization does not replace the measured resonator frequency.

Task update targets

Task update targets are fixed and have the same meaning in standalone runs and workflows. Publication requires both enabled output persistence and successful validation.

ExperimentFrequency output to publishEffect
Qubit Spectroscopy, including 2D and 1D frequency estimationcoarse_qubit_frequencyUpdate the exploration estimate
CheckControlAmplitude, when its frequency fit succeedscoarse_qubit_frequencyRefine the exploration estimate
Chevronqubit_frequencyUpdate the calibrated qubit frequency
Ramseyqubit_frequencyUpdate the calibrated qubit frequency
Resonator Spectroscopyresonator_frequencyUpdate the measured resonator frequency
CKP, for its resonator-frequency estimateresonator_frequencyUpdate the measured resonator frequency
Optimal Readout Frequencyoptimal_readout_frequencyUpdate the readout optimization result

The spectroscopy rows describe measurement methods, not a requirement to introduce a task class for each method. The CKP mapping assumes that the relevant output estimates the resonator frequency; its concrete task and output contract require verification before implementation.

CheckCoarseReadoutParams extracts a result named optimal_readout_frequency, selected by maximizing the Rabi IQ response range across readout frequencies and amplitudes. This objective is different from directly optimizing readout fidelity. If this output is treated as readout optimization, its destination is optimal_readout_frequency, and the UI must disclose replacement of a value from another optimization task. Confirm that this shared destination is appropriate before changing its output name.

Update targets are specified per output, not by classifying the entire task as exploration or calibration. Amplitude and other outputs retain their own declared destinations. In particular, Resonator Spectroscopy also produces readout_amplitude; maintaining the readout frequency does not imply that every readout setting remains unchanged.

Replacement within the same parameter

There is no automatic ranking of task names or calibration stages. With publication enabled, a validated Chevron result can replace a Ramsey result in qubit_frequency. Likewise, a validated Resonator Spectroscopy result can replace a CKP result in resonator_frequency when that mapping is confirmed.

The execution UI explains the current source and the replacement before the run. Operators can choose measurement-only execution. This policy protects calibrated qubit frequencies from exploration updates; it does not guarantee that every accepted recalibration improves precision or device performance.

Runtime frequency resolution

For an ordinary execution without explicit overrides, resolve the default drive frequencies as follows:

text
control_frequency = qubit_frequency

readout_frequency = optimal_readout_frequency, if set
                    otherwise resonator_frequency

“Set” means a usable, accepted value, not truthiness of a number. Invalid values fail validation rather than silently selecting a fallback. If a required frequency cannot be resolved, fail with the missing parameter identified.

Do not automatically copy or fall back from coarse_qubit_frequency into qubit_frequency or the default control drive frequency. Exploration and Chevron tasks explicitly consume the exploration value where required. Experiments investigating the resonator explicitly consume resonator_frequency for their measurement needs; this is separate from resolving a default readout drive frequency.

Permitted per-execution overrides remain explicit inputs. An arbitrary drive override does not become a measured qubit or resonator frequency. Record the actual resolved drive frequencies, their sources, and overrides in execution history even though the drive frequencies are not independent persistent calibration parameters.

Snapshot re-execution uses the recorded effective inputs and permitted overrides, not current database values or current YAML defaults. Preserve enough source information in the snapshot to reproduce the frequency selection.

Bringup and repeated measurements

Bringup uses the normal task destinations rather than a separate calibration-stage reset or an all-workflow publication checkpoint:

  1. Qubit Spectroscopy updates coarse_qubit_frequency after validation.
  2. CheckControlAmplitude consumes and refines the exploration parameters.
  3. Chevron consumes the exploration parameters and publishes qubit_frequency after validation.
  4. Subsequent calibration tasks consume the updated calibrated values.

Validated outputs are published as the workflow progresses when persistence is enabled. A failure in a later task does not roll back earlier accepted outputs. Report the outputs already updated and the tasks that did not complete. Failed outputs must not become the normal inputs of downstream tasks.

An initial bringup can reach Chevron without an existing qubit_frequency, using its explicit exploration inputs. Repeating only Qubit Spectroscopy on a calibrated device updates the exploration value while retaining the calibrated value. Repeating bringup can replace the calibrated value when Chevron succeeds; the start screen identifies bringup as a recalibration operation.

Readout transitions

Operation with publication enabled and validation passedresonator_frequencyoptimal_readout_frequencyResolved readout_frequency
First Resonator SpectroscopyUpdateUnsetNew resonator frequency
Readout frequency optimizationRetainUpdateNew optimized frequency
Repeat Resonator Spectroscopy after optimizationUpdateRetainRetained optimized frequency
Explicitly clear the accepted optimization valueRetainUnsetCurrent resonator frequency

Remeasuring the resonator does not automatically erase readout optimization. Changes to cooldown or readout conditions can require optimization to be repeated. Provide an explicit action to clear the accepted optimization value and use the resonator frequency again. Retain the old measurement in history and record the clearing action.

Frequency selection alone does not establish that an old optimization remains suitable after other readout settings change. Show other updated outputs, such as readout amplitude, alongside the frequency behavior. Do not claim that all readout settings are preserved when only the frequency is retained.

Coarse readout scan settings

CheckCoarseReadoutParams declares the calibration inputs used by both the readout scan and its underlying Rabi measurements:

Input parameterPurposeResolution
qubit_frequencyQubit drive frequency for the Rabi measurementsRequired database value
control_amplitudeControl pulse amplitude for the Rabi measurementsRequired database value
readout_frequencyCenter frequency of the readout scanRequired database value
readout_amplitudeReference amplitude of the readout scanRequired database value

All four calibration inputs are recorded for snapshot re-execution. Explicit input overrides define the values for that run; the scan does not independently reload them from Qubex YAML. Drive frequency and amplitude overrides are scoped to the scan and restored on success or failure. The session-scoped readout_duration is recorded as a Run parameter. The Qubex contrib helper has no readout-duration argument, so a task-local adapter forwards that value to each underlying Rabi call without replacing methods on the shared Experiment instance.

Run parameterDefaultMeaning
readout_durationQubex session valueReadout pulse duration in ns for all underlying Rabi measurements
detuning_range(-0.015, 0.015, 13)Frequency offsets in GHz, as (start, stop, number of points)
readout_amplitude_ratio_range(0.8, 1.2, 5)Multipliers of the reference amplitude, as (start, stop, number of points)
time_range(0, 101, 4)Rabi drive durations in ns, as (start, exclusive stop, step)
shots1024Shots per Rabi drive duration at each readout point
intervalExisting Qubex defaultShot interval in ns

The default scan includes the current frequency and amplitude: 13 frequency points over ±15 MHz and 5 amplitude points over ±20%. Amplitudes are capped at 1 and duplicate points are removed, so a reference near the upper bound can produce fewer than 5 amplitude points. The reference amplitude must be finite and within (0, 1]; invalid references fail rather than being replaced or clipped.

Normally this performs 65 Rabi traces, compared with the previous 147-point grid using ±25 MHz and amplitudes from zero to the configured value. The shot count is halved from 2048 to 1024, while the default Rabi time window and shot interval are retained. These are adjustable starting settings, not a claim of equivalent measurement quality. Operators can widen the scan, change the Rabi time window, or restore 2048 shots using run parameters.

The progress plan uses the effective frequency count multiplied by the effective amplitude count, after clipping and duplicate removal. The UI shows all-sweep progress from the first Rabi trace: normally 65 sweeps, or 39 with a reference amplitude of 1 and the default ranges. The displayed count is completed sweeps out of the total; the percentage also includes progress within the active sweep. After the first sweep completes, the remaining-time estimate extrapolates the total elapsed measurement time across the planned sweeps. It includes time between sweeps and remains an estimate, since measurement speed can vary. The UI hides the inner Rabi time-point count and per-sweep ETA for this search.

Older snapshots without the newly declared reference inputs cannot silently use current settings; handle missing snapshot inputs according to Parameter Resolution.

Coarse readout R² validation

CheckCoarseReadoutParams validates the Rabi fit at the selected readout frequency and amplitude before accepting either output. In the installed Qubex revision pinned by uv.lock, rabi_results contains the scan results in amplitude-major order; each result exposes rabi_params[qubit_label].r2. Use this stored R² without refitting or averaging the scan.

The default threshold is shared with QDash's CheckRabi: 0.6. As with the existing task executor, a value must be strictly greater than the threshold to pass. Missing, non-finite, or greater-than-one R² values also fail validation. Selection metadata that cannot identify the matching scan point is a validation failure, not permission to skip the check.

Do not require every scan point to pass: exploration can include low readout amplitudes and other settings with poor responses. Conversely, a good fit at an unrelated point must not authorize publishing a noisy selected point. Do not silently choose a different candidate when the selected point fails.

Preserve the heatmap and available raw scan data for inspection, mark the task failed, and publish neither frequency nor amplitude on rejection. The selected R² is recorded as task quality metadata when finite. This gate is implemented in the task; it does not establish that the response maximum is the fidelity optimum or guarantee rejection of every possible noise realization.

YAML export

YAML files represent accepted parameter values and the default drive configuration derived from them. They do not export every new measurement automatically.

YAML fileExported value
coarse_qubit_frequency.yamlAccepted exploration qubit frequency
qubit_frequency.yamlAccepted calibrated qubit frequency
control_frequency.yamlAccepted qubit_frequency
resonator_frequency.yamlAccepted measured resonator frequency
optimal_readout_frequency.yamlAccepted readout optimization result, when set
readout_frequency.yamlResult of the runtime readout selection rule

coarse_qubit_frequency.yaml and optimal_readout_frequency.yaml preserve and share their respective values even if Qubex does not consume them directly. Verify compatibility with the deployed Qubex parameter loader before placing additional files in its parameter directory.

On accepted exploration updates, update only the exploration frequency file among the qubit frequency files. On accepted Chevron or Ramsey updates, update both qubit_frequency.yaml and control_frequency.yaml.

On accepted resonator updates, update resonator_frequency.yaml and recompute readout_frequency from the full accepted state. An existing optimized frequency remains selected. On accepted optimization updates, update optimal_readout_frequency.yaml and the derived readout_frequency.yaml. Do not overwrite resonator_frequency.yaml with an optimized drive frequency.

Clearing an optimization value must remove or unset the affected target's exported optimization value and rewrite its readout drive frequency using the resonator fallback. Do not leave a stale optimized value in either file or remove other targets' values. If the fallback cannot be resolved, report missing configuration rather than silently retaining the old drive frequency.

Use one frequency selection rule for execution and YAML generation. A static mapping that copies each changed source into readout_frequency.yaml cannot implement the required precedence. Publication must keep the accepted database state, normal downstream inputs, and exported defaults consistent; export failures must be visible rather than reported as fully applied updates.

QDash-generated defaults are managed outputs. Independent edits to control_frequency.yaml or readout_frequency.yaml are not QDash calibration inputs. For standalone Qubex experiments requiring different defaults, use separate configuration or explicit execution overrides. QDash cannot guarantee identical behavior while ignoring independent edits to shared drive-frequency files.

Execution UX

Execution intent

Expose the execution purpose and update destinations in user-facing language rather than relying on an ambiguous “Update params” switch.

ActionInput baselineDefault publication behavior
Measure onlyCurrent calibration state, plus permitted overridesHistory only
Calibrate and applyCurrent calibration state, plus permitted overridesPublish validated outputs to declared destinations
Repeat with the same conditionsRecorded snapshot, plus permitted overridesHistory only
BringupCurrent calibration state and outputs from preceding stepsPublish validated outputs as tasks complete

Keep measurement-only execution as the initial selection for Tasks quick runs. A new run using current settings and a snapshot re-execution must be distinguishable in navigation and labels. Publication is separate from input selection and from bypassing validation.

Before execution

Show the update destinations and their current source task near the execution action. For spectroscopy that publishes exploration results, use explanatory text such as:

Updates the exploration frequency (coarse_qubit_frequency). The calibrated qubit frequency is retained; its current source is Chevron.

For Chevron replacing a Ramsey calibration:

Updates qubit_frequency. The current value comes from Ramsey. A validated result will replace it with this Chevron measurement.

For Resonator Spectroscopy after readout optimization:

Updates resonator_frequency. Readout continues to use the accepted Optimal Readout Frequency result.

List other updated outputs separately. Generate the source descriptions from the selected target's actual calibration metadata. Measurement-only runs must instead state that results are recorded without applying parameter updates. Keep the measurement-only choice near the action; routine execution does not require another confirmation dialog.

After execution

Display execution outcome and publication outcome separately. For each output, show the measured value, whether it was applied, the previous accepted value where applicable, and the reason when it was not applied. Show the retained operating frequency and its source when an exploration result was updated.

Use explicit outcomes such as “Applied,” “History only,” and “Not applied: validation failed.” A successful measurement alone does not mean the calibration database or YAML was updated. YAML synchronization failures must remain distinguishable from measurement failures.

Implementation gaps

The policy requires changes beyond adding YAML mappings. At the time this policy was added:

  • Qubit Spectroscopy and CheckControlAmplitude already output coarse_qubit_frequency; Chevron and Ramsey already output qubit_frequency.
  • Resonator Spectroscopy outputs its resonator estimate as readout_frequency. Optimal Readout Frequency and CheckCoarseReadoutParams also output readout_frequency, conflating the measured resonator frequency and readout optimization.
  • config/app/workflow.yaml maps resonator_frequency to readout_frequency.yaml. It already exports qubit_frequency to both qubit and control frequency files, but does not map the coarse qubit or optimal readout frequency files. The publication allowlist also needs to include the new exports.
  • Many task inputs request readout_frequency directly from the database. These inputs, snapshot handling, and the Qubex adapter need a shared resolution path before the independent database parameter can be retired.
  • The executor clears rejected outputs before calling BackendSaver for postprocess and R² validation failures. BackendSaver itself writes non-empty calibration outputs before its backend-success gate. Preserve the executor's rejection path when implementing this policy; setting only the backend-success flag is insufficient to protect database values.
  • persist_output_parameters controls output publication. In the single-task flow, update_params is passed as force_update_params, which can bypass the backend quality gate. The re-execution UI labels it “Update backend params”; this must not serve as the ordinary “Calibrate and apply” control.

Existing readout_frequency records may originate from spectroscopy or optimization. Migration must use task provenance where available rather than copying every legacy value into both new parameters. Ambiguous records require an explicit classification or remeasurement. Preserve historical records and their original meanings, including snapshots used for re-execution.