Cross-Resonance Tasks
CheckCrossResonance measures CR parameters; CreateZX90 fits the amplitude for a ZX90 gate. Qubex measures and fits; QDash validates settings and accepted outputs.
Run parameters
Set these in a TaskCall.run entry:
| Parameter | Constraint |
|---|---|
ramptime | Finite and positive |
adiabatic_safe_factor, max_amplitude | In (0, 1] |
duration (ZX90) | Positive and at least twice the ramp time |
amplitude_range (ZX90) | Increasing, within the cap, at least four points |
rotary_multiple (ZX90) | Positive |
duration_buffer (ZX90) | At least 1 |
An input CR duration of zero lets Qubex derive it; a nonzero duration must cover both ramps. Stored ramp time must match the requested value. No ramp time is universally safe for all pairs.
Amplitude caps
The cap combines max_amplitude with Qubex's adiabatic bound. CheckCrossResonance publishes the bounds it used; CreateZX90 inherits them or uses legacy defaults of 0.75 and 1.0. A run may tighten inherited bounds but cannot loosen them. To change the accepted bound, rerun characterization or explicitly override its input.
CR duration
With no explicit duration/range, Qubex can lengthen duration once when its rough scan finds no root. An explicit duration stays fixed and a missing root fails. An explicit amplitude range skips the rough scan and its retry. Compare published duration with input when using defaults.
Accepting a result
CreateZX90 requires a real fitted root matching the stored amplitude, inside the scanned range and cap. Requested duration and ramp time must match. Amplitude 1.0 alone is not proof of a root; Qubex also uses that value when fitting fails. CheckCrossResonance requires finite outputs.
Rejected fits retain figures and restore the CR note entry captured immediately before the Qubex call. They do not undo inputs synchronized earlier by preprocessing. Only accepted results save the note file.
Fresh results
Both tasks require a newer note timestamp after the Qubex call, preventing an old entry from masquerading as a result. Qubex timestamps have one-second resolution, so this proves a write in the run, not that the last tomography iteration replaced an earlier same-second iteration.
Measurement-only runs
update_calibration_parameters=False blocks QDash database publication, but accepted CR results still update and save the Qubex note. It does not make these tasks non-destructive.
Testing
tests/qdash/workflow/calibtasks/qubex/test_cr_zx90_controls.py covers bounds, rejected-fit restoration, and freshness with real calibration-note behavior and mocked measurements. It does not establish hardware fidelity or safe pulse choices.
Implementation files
Under src/qdash/workflow/calibtasks/qubex/two_qubit/: cr_controls.py, check_cross_resonance.py, and create_zx90.py.